📝 Event: 2026-02-24_installation_laser_baseline_qc
📊 Report Overview
Event Metadata
Record Type
Qc Session
Qc Session
Reason / Suite
Post_installation_baseline
Post_installation_baseline
Operator
Installation engineer
Installation engineer
Instrument
scope-andor-bc43QC Summary & Evaluation
Summary: Baseline laser power check completed during BC43 installation acceptance.
🔬 Tests Performed
| Test Type | Details |
|---|---|
| Laser Power | Single-point baseline measurement for each installed laser line. |
📏 Computed Metrics
| Metric ID | Value |
|---|---|
laser.at_obj.405.100pct.power_mw |
18.0 |
laser.at_obj.488.100pct.power_mw |
12.0 |
laser.at_obj.561.100pct.power_mw |
13.0 |
laser.at_obj.638.100pct.power_mw |
16.0 |
Additional Notes
Laser power measured with calibrated laser slides provided with the instrument package. Values captured as initial reference for future trending.
📄 Raw YAML Ledger
View 2026-02-24_installation_laser_baseline_qc.yaml
schema_version: 1
record_type: qc_session
qc_id: "qc_scope-andor-bc43_20260224T110000Z_laser_baseline"
microscope: "scope-andor-bc43"
started_utc: "2026-02-24T11:00:00Z"
ended_utc: "2026-02-24T11:20:00Z"
performed_by: "Installation engineer"
reason: "post_installation_baseline"
summary: "Baseline laser power check completed during BC43 installation acceptance."
notes: >
Laser power measured with calibrated laser slides provided with the instrument package.
Values captured as initial reference for future trending.
performed:
- qc_type: "laser_power"
details: "Single-point baseline measurement for each installed laser line."
artifacts: []
inputs_human:
- metric_id: "visual.optical_path_integrity"
value: "verified"
unit: ""
details: "Objectives and optical pathway integrity checked during installation acceptance."
laser_inputs_human:
measurement_position: "at_objective"
measurement_position_details: "Measured using calibrated laser slides supplied with the facility."
power_meter:
model: "calibrated laser slides"
serial: ""
calibration_due: ""
integration_time_s: 1.0
notes: "Installation acceptance baseline."
linearity_series:
- laser: "405"
setpoint_units: "percent"
power_units: "mW"
points:
- setpoint: 100
power: 18.0
details: "Baseline measured during installation acceptance."
- laser: "488"
setpoint_units: "percent"
power_units: "mW"
points:
- setpoint: 100
power: 12.0
details: "Baseline measured during installation acceptance."
- laser: "561"
setpoint_units: "percent"
power_units: "mW"
points:
- setpoint: 100
power: 13.0
details: "Baseline measured during installation acceptance."
- laser: "638"
setpoint_units: "percent"
power_units: "mW"
points:
- setpoint: 100
power: 16.0
details: "Baseline measured during installation acceptance."
stability_series: []
computed_provenance:
pipeline: "manual_entry"
version: "1"
run_utc: "2026-02-24T11:30:00Z"
metrics_computed:
- metric_id: "laser.at_obj.405.100pct.power_mw"
value: 18.0
unit: "mW"
details: "Baseline max power at objective measured during installation."
- metric_id: "laser.at_obj.488.100pct.power_mw"
value: 12.0
unit: "mW"
details: "Baseline max power at objective measured during installation."
- metric_id: "laser.at_obj.561.100pct.power_mw"
value: 13.0
unit: "mW"
details: "Baseline max power at objective measured during installation."
- metric_id: "laser.at_obj.638.100pct.power_mw"
value: 16.0
unit: "mW"
details: "Baseline max power at objective measured during installation."
artifacts: []