📝 Event: 2025-02-11_preventive_maintenance
📊 Report Overview
Event Metadata
Record Type
Maintenance Event
Maintenance Event
Reason / Suite
Scheduled
Scheduled
Service Provider
vendor
vendor
Instrument
scope-testx1Maintenance Action
Reason Details:
Annual preventive maintenance visit after high usage in Q4. Users had noticed mild focus drift during long 60x live-cell runs.
Annual preventive maintenance visit after high usage in Q4. Users had noticed mild focus drift during long 60x live-cell runs.
Action Details:
Completed vendor PM checklist, cleaned objective turret contacts, checked stage cable strain relief, and adjusted incubation chamber fan profile to reduce vibration transfer.
Completed vendor PM checklist, cleaned objective turret contacts, checked stage cable strain relief, and adjusted incubation chamber fan profile to reduce vibration transfer.
Status After Fix:
IN_SERVICE
Additional Notes
Drift reduced after fan profile adjustment. No component replacement needed.
📄 Raw YAML Ledger
View 2025-02-11_preventive_maintenance.yaml
schema_version: 1
record_type: maintenance_event
maintenance_id: "maint_scope-testx1_20250211_preventive_maintenance"
microscope: "scope-testx1"
service_provider: "vendor"
company: "Nordic Imaging Service Oy"
contact: "Field Engineer: Laura Niemi (laura.niemi@nordicimaging.fi)"
reference: "NIS-2025-0211-044"
started_utc: "2025-02-11T07:45:00Z"
ended_utc: "2025-02-11T11:20:00Z"
reason: "scheduled"
reason_details: >
Annual preventive maintenance visit after high usage in Q4.
Users had noticed mild focus drift during long 60x live-cell runs.
action: "service"
action_details: >
Completed vendor PM checklist, cleaned objective turret contacts,
checked stage cable strain relief, and adjusted incubation chamber
fan profile to reduce vibration transfer.
microscope_status_after: "in_service"
downtime_hours: 3.6
related_qc:
- "qc_scope-testx1_20250210T0910Z_psf_v1"
- "qc_scope-testx1_20250212T0835Z_psf_v1"
attachments:
- "maintenance/events/scope-testx1/2025/attachments/2025-02-11_pm_report.pdf"
tags: ["scheduled", "focus", "vibration", "vendor"]
followup: >
Re-check autofocus repeatability after two weeks of incubated imaging.
next_due_date: "2026-02-01"
notes: >
Drift reduced after fan profile adjustment. No component replacement needed.