Maintenance event · 2017-05-17
Event overview
Event metadata
Record type
Maintenance event
Maintenance event
Reason / suite
Scheduled
Scheduled
Service provider
Vendor
Vendor
Record ID
2017-05-17_preventative_maintenanceMaintenance action
Reason details:
scheduled
scheduled
Action details:
Preventative Maintenance Visit. Addressed MPU-AC issue where the main switch was broken. The main switch at the MPU was replaced and functionality restored. The Evolve camera was found misaligned and was realigned/calibrated using RapidCal. Software SlideBook was upgraded to 6.0.12 and NIDAQ critical updates were applied. Kohler, Vector, and FRAP light paths were realigned. System surfaces and filters were cleaned.
Preventative Maintenance Visit. Addressed MPU-AC issue where the main switch was broken. The main switch at the MPU was replaced and functionality restored. The Evolve camera was found misaligned and was realigned/calibrated using RapidCal. Software SlideBook was upgraded to 6.0.12 and NIDAQ critical updates were applied. Kohler, Vector, and FRAP light paths were realigned. System surfaces and filters were cleaned.
Raw YAML record
View 2017-05-17_preventative_maintenance.yaml
record_type: maintenance_event
microscope: scope-3i-csu-w1-spinning-disk
maintenance_id: maint-2017-05-17-pm
started_utc: '2017-05-17T09:00:00Z'
completed_utc: '2017-05-17T17:00:00Z'
status: completed
reason: scheduled
action: replace
action_details: 'Preventative Maintenance Visit. Addressed MPU-AC issue where the
main switch was broken. The main switch at the MPU was replaced and functionality
restored. The Evolve camera was found misaligned and was realigned/calibrated using
RapidCal. Software SlideBook was upgraded to 6.0.12 and NIDAQ critical updates
were applied. Kohler, Vector, and FRAP light paths were realigned. System surfaces
and filters were cleaned.
'
service_provider: vendor
company: 3i (Intelligent Imaging Innovations)
contact: 3i Field Service Engineer
status_notes: Fully functional