Skip to content
← Advanced Imaging Core Facility at Turku Bioscience Centre

Leica DM RE

About this microscope

Here you’ll find the hardware and capabilities currently recorded for this microscope. If you are planning an experiment, AIC Turku staff can help with access, training, and configuration choices.

Current status: 🟢 Online

Operational

QC and maintenance history Open Virtual Microscope

Leica DM RE microscope
Manufacturer
Leica Microsystems
Model
Leica DM RE
Stand
Upright
Eyepieces
Trinocular
Location
BioCity, 2nd floor, room 2076
Purchased
—
Funding
—
Capabilities
These are the capabilities recorded for this microscope. Open the optical-route details below to see how the hardware is connected.
Transmitted brightfieldDICPhase contrast
Methods and the light path that records them
Each method below is explicitly associated in the instrument record with the physical light path that implements it. An imaging mode or contrast method that appears above but not here has no recorded method-to-path mapping yet.
  • Transmitted light: Transmitted brightfield, DIC, Phase contrast


Software

No acquisition/control software — manual or standalone system.


Hardware

Light sources

  • Halogen Lamp Leica 12V 100W halogen bulb
    Leica
    Type: Halogen Lamp
    Role: Transmitted Illumination
    Power: 100 W
    Notes: Transmitted light source for TL and DIC. Broadband illumination.

Detectors

  • Unknown Unknown Camera
    Charge-Coupled Device
    Type: Charge-Coupled Device
    Notes: Camera port in microscope body for optional detector.

Objectives

View this microscope's objectives in the catalogue

  • HC PL FLUOTAR 10x/0.30 PH1 AIR
    Leica
    Model: HC PL FLUOTAR 10x/0.30 PH1
    Magnification / NA: 10x/0.3
    Immersion: Air
    Correction: Plan Fluorite
    Working distance: 11.0 mm
    Product code: 506507
    Installed: Yes
    Specialties: phase, dic
    Notes: The facility Leica DMRE getting-started record identifies the installed 10x as article 506507 with PH1. Independent exact-code specifications give an 11.0 mm working distance.

  • HC PL FLUOTAR 20x/0.50 PH2 AIR
    Leica
    Model: HC PL FLUOTAR 20x/0.50 PH2
    Magnification / NA: 20x/0.5
    Immersion: Air
    Correction: Plan Fluorite
    Working distance: 1.15 mm
    Product code: 506506
    AFC compatible: No
    Installed: Yes
    Specialties: phase, dic
    Notes: Working distance verified against an exact 506506 facility technical specification.

  • PL FLUOTAR 40x/0.70 PH2 AIR
    Leica
    Model: PL FLUOTAR 40x/0.70 PH2
    Magnification / NA: 40x/0.7
    Immersion: Air
    Correction: Plan Fluorite
    Product code: 506014
    AFC compatible: No
    Installed: Yes
    Specialties: phase, dic
    Notes: The facility DMRE getting-started record labels this installed 506014 objective as Ph3, but independent exact-code listings identify 506014 as Ph2, and facility staff confirmed Ph2 on 28 September 2026. The getting-started record is in error on this point. Working distance remains unknown.

  • PL APO 63x/1.32 OIL PH3 OIL
    Leica
    Model: PL APO 63x/1.32 OIL PH3
    Magnification / NA: 63x/1.32
    Immersion: Oil
    Correction: Plan Apochromat
    Working distance: 0.07 mm
    Product code: 506082
    AFC compatible: No
    Installed: Yes
    Specialties: phase, dic
    Notes: Working distance is 0.07 mm; 0.17 mm is the cover-glass specification and must not be used as WD.

  • PL FLUOTAR 100x/1.30 OIL PH3 OIL
    Leica
    Model: PL FLUOTAR 100x/1.30 OIL PH3
    Magnification / NA: 100x/1.3
    Immersion: Oil
    Correction: Plan Fluorite
    Product code: 506017
    AFC compatible: No
    Installed: Yes
    Specialties: phase, dic


Latest QC measurements

Laser performance

No recent laser-performance measurements are recorded.

Optical resolution (PSF and alignment)

No recent PSF or chromatic-shift measurements are recorded for these objectives.

Objectives without recent measurements
  • HC PL FLUOTAR 10x/0.30 PH1 AIR (10x_air)

  • HC PL FLUOTAR 20x/0.50 PH2 AIR (20x_air)

  • PL FLUOTAR 40x/0.70 PH2 AIR (40x_air)

  • PL APO 63x/1.32 OIL PH3 OIL (63x_oil)

  • PL FLUOTAR 100x/1.30 OIL PH3 OIL (100x_oil)

Information still to confirm

Some details in the facility record are incomplete. If any of these matter for your experiment, please check them with AIC Turku staff.

  • Details needed for this configuration:

  • Detector pixel pitch (um) (hardware.detectors[].pixel_pitch_um)