Abberior STED
About this microscope
Here you’ll find the hardware and capabilities currently recorded for this microscope. If you are planning an experiment, AIC Turku staff can help with access, training, and configuration choices.
- Point-scanning confocal: Confocal point scanning, STED, RESOLFT
Software
- Abberior Imspector
Acquisition
Role: Acquisition
Version: Not documented
Hardware
Light sources
-
485 nm Laser PicoQuant LDH-D-C-485
PicoQuant
Type: Laser
Role: Excitation
Timing mode: Pulsed
Pulse width:80 ps
Repetition rate:40 MHz
Wavelength:485 nm
Notes: Pulsed excitation source for the 488 nm channel. -
561 nm Laser PicoQuant PDL-T 561
PicoQuant
Type: Laser
Role: Excitation
Timing mode: Pulsed
Pulse width:80 ps
Repetition rate:40 MHz
Wavelength:561 nm
Notes: Pulsed excitation source for the 561 nm channel. -
640 nm Laser PicoQuant LDH-D-C-640
PicoQuant
Type: Laser
Role: Excitation
Timing mode: Pulsed
Pulse width:80 ps
Repetition rate:40 MHz
Wavelength:640 nm
Notes: Pulsed excitation source for the 640 nm channel. -
580 nm Laser MPB Communications VFL-P-1000-580
MPB Communications
Type: Laser
Role: Depletion
Timing mode: Continuous Wave
Depletion targets:488 nm
Wavelength:580 nm
Notes: Continuous-wave STED depletion source for the 488 nm excitation channel. -
775 nm Laser OneFive / NKT Photonics Katana HP 775
OneFive / NKT Photonics
Type: Laser
Role: Depletion
Timing mode: Pulsed
Pulse width:600 ps
Repetition rate:40 MHz
Depletion targets:561 nm,640 nm
Wavelength:775 nm
Notes: Pulsed STED depletion source for the 561 nm and 640 nm excitation channels.
Scanner
- Galvanometric Scanner
Type: Galvanometric Scanner
Notes: STED point scanning configuration.
Detectors
-
Hamamatsu Photonics Photomultiplier Tube
Photomultiplier Tube
Type: Photomultiplier Tube
Notes: General confocal detection channel. -
Excelitas Technologies SPCM-AQRH-14-TR
Avalanche Photodiode
Type: Avalanche Photodiode
Supports time gating: Yes
Default gating delay:1 ns
Default gate width:7 ns
Notes: Two APD detectors; timing resolution optimized for time-gated CW-STED. -
Generic Monochrome CCD camera
Charge-Coupled Device
Type: Charge-Coupled Device
Pixel pitch:4.65 µm
Sensor format:1280 x 960
Notes: 1/2-inch format camera.
Objectives
View this microscope's objectives in the catalogue
-
LMPlanFL N 20x/0.40 AIR
Olympus
Model: LMPlanFL N 20x/0.40
Magnification / NA:20x/0.4
Immersion: Air
Correction: Plan Fluorite
Working distance:12.0 mm
Product code:1-U2N345
Installed: Yes -
UPlanSApo 60x/1.2 W WATER
Olympus
Model: UPlanSApo 60x/1.2 W
Magnification / NA:60x/1.2
Immersion: Water
Correction: Plan Apochromat
Working distance:0.28 mm
Product code:N6432600
Installed: Yes -
UPlanSApo 100x/1.40 Oil OIL
Olympus
Model: UPlanSApo 100x/1.40 Oil
Magnification / NA:100x/1.4
Immersion: Oil
Correction: Plan Apochromat
Working distance:0.13 mm
Product code:1-U2B836
Installed: Yes
Optical modulators
- Spatial Light Modulator
Abberior
Type: Spatial Light Modulator
Supported phase masks:Vortex,Bottle,3D-STED
Notes: Used for Easy3D STED phase-pattern generation and adaptive optics workflows.
Illumination control
- RESCue STED
Adaptive illumination
Method: RESCue STED
Default enabled: No
Notes: Real-time adaptive STED decision logic used to reduce photobleaching and light dose during live-cell super-resolution imaging.
Optical-path elements
- Emission Filter Wheel
Emission
Element type: Filter Wheel
Pos 1: Bandpass (FF01-525/25) — emission 525/25 nm
Pos 2: Bandpass — emission 615/10 nm
Pos 3: Bandpass (ET685/35) — emission 685/35 nm
Latest QC measurements
Laser performance
No recent laser-performance measurements are recorded.
Optical resolution (PSF and alignment)
No recent PSF or chromatic-shift measurements are recorded for these objectives.
Objectives without recent measurements
-
LMPlanFL N 20x/0.40 AIR (
20x_air) -
UPlanSApo 60x/1.2 W WATER (
60x_water) -
UPlanSApo 100x/1.40 Oil OIL (
100x_oil)
Information still to confirm
Some details in the facility record are incomplete. If any of these matter for your experiment, please check them with AIC Turku staff.
-
Details needed for this configuration:
-
Software version (
software[].version) -
Detector peak quantum efficiency (%) (
hardware.detectors[].qe_peak_pct)