EVOS fl
About this microscope
Here you’ll find the hardware and capabilities currently recorded for this microscope. If you are planning an experiment, AIC Turku staff can help with access, training, and configuration choices.
- Widefield fluorescence: Widefield fluorescence
- Transmitted light: Transmitted brightfield, Phase contrast
Software
- On-board EVOS interface
Acquisition
Role: Acquisition
Version: Not documented
Notes: no external software, USB storage
Hardware
Light sources
-
LED Thermo Fisher / AMG Transmitted Light LED
Thermo Fisher / AMG
Type: LED
Notes: Broad spectrum overhead LED used for brightfield and phase contrast. -
357 nm LED Thermo Fisher / AMG EVOS DAPI Light Cube LED
Thermo Fisher / AMG
Type: LED
Wavelength:357 nm
Notes: Integrated LED inside the DAPI Light Cube -
445 nm LED Thermo Fisher / AMG EVOS CFP Light Cube LED
Thermo Fisher / AMG
Type: LED
Wavelength:445 nm
Notes: Integrated LED inside the CFP Light Cube -
470 nm LED Thermo Fisher / AMG EVOS GFP Light Cube LED
Thermo Fisher / AMG
Type: LED
Wavelength:470 nm
Notes: Integrated LED inside the GFP Light Cube -
531 nm LED Thermo Fisher / AMG EVOS RFP Light Cube LED
Thermo Fisher / AMG
Type: LED
Wavelength:531 nm
Notes: Integrated LED inside the RFP Light Cube -
628 nm LED Thermo Fisher / AMG EVOS Cy5 Light Cube LED
Thermo Fisher / AMG
Type: LED
Wavelength:628 nm
Notes: Integrated LED inside the Cy5 Light Cube
Detectors
-
AMG (Thermo Fisher) Sony ICX285AQ Color CCD
Charge-Coupled Device
Type: Charge-Coupled Device
Pixel pitch:6.45 µm
Notes: Integrated color camera for brightfield (1.4 Megapixels). -
AMG (Thermo Fisher) AMF-4302 (Sony ICX285AL Monochrome CCD)
Charge-Coupled Device
Type: Charge-Coupled Device
Pixel pitch:6.45 µm
Notes: Integrated monochrome camera for fluorescence (1.4 Megapixels).
Objectives
View this microscope's objectives in the catalogue
-
U Plan FL N 4x/0.13 PhP AIR
Olympus
Model: U Plan FL N 4x/0.13 PhP
Magnification / NA:4x/0.13
Immersion: Air
Correction: Fluorite
Working distance:16.4 mm
AFC compatible: No
Installed: Yes
Specialties: phase
Notes: PhP -
Plan Fluor 10x/0.3 AIR
AMG (Thermo Fisher)
Model: Plan Fluor 10x/0.3
Magnification / NA:10x/0.3
Immersion: Air
Correction: Plan Fluorite
Working distance:8.3 mm
Product code:AMG-AMEP 4623
AFC compatible: No
Installed: Yes
Notes: Fluorite (FL) -
Plan Fluor 20x/0.45 AIR
AMG (Thermo Fisher)
Model: Plan Fluor 20x/0.45
Magnification / NA:20x/0.45
Immersion: Air
Correction: Plan Fluorite
Working distance:7.1 mm
Product code:AMG-AMEP 4624
AFC compatible: No
Installed: Yes
Notes: Fluorite (FL) -
Plan Fluor 40x/0.65 AIR
AMG (Thermo Fisher)
Model: Plan Fluor 40x/0.65
Magnification / NA:40x/0.65
Immersion: Air
Correction: Plan Fluorite
Working distance:2.8 mm
Product code:AMG-AMEP 4625
AFC compatible: No
Installed: Yes
Notes: Fluorite (FL)
Optical-path elements
- Light Cube Turret
Cube
Element type: Turret
Details: Filter sets must be changed manually. Only AIC personnel are allowed to change filter cubes.
Pos 1: DAPI (ZP-EPI-9001) — emission 447/60 nm — Included with base unit. Excitation: 357/44 nm. Dichroic Cutoff: 400 nm.
Pos 2: CFP (AMEP-4653) — emission 510/42 nm — Excitation: 445/45 nm. Dichroic Cutoff: 480 nm.
Pos 3: GFP/Alexa 488 (ZP-EPI-9002) — emission 510/42 nm — Included with base unit. Excitation: 470/22 nm. Dichroic Cutoff: 495 nm.
Pos 4: RFP/Alexa 555 (ZP-EPI-9003) — emission 593/40 nm — Included with base unit. Excitation: 531/40 nm. Dichroic Cutoff: 560 nm.
Pos 5: Cy5/Alexa 647 (AMEP-4656) — emission 692/40 nm — Excitation: 628/40 nm. Dichroic Cutoff: 660 nm.
Latest QC measurements
Laser performance
No recent laser-performance measurements are recorded.
Optical resolution (PSF and alignment)
No recent PSF or chromatic-shift measurements are recorded for these objectives.
Objectives without recent measurements
-
U Plan FL N 4x/0.13 PhP AIR (
4x_air) -
Plan Fluor 10x/0.3 AIR (
10x_air) -
Plan Fluor 20x/0.45 AIR (
20x_air) -
Plan Fluor 40x/0.65 AIR (
40x_air)
Information still to confirm
Some details in the facility record are incomplete. If any of these matter for your experiment, please check them with AIC Turku staff.
-
Details needed for this configuration:
-
Software version (
software[].version)