System Health Overview
This page lists instruments currently in a degraded state (Warning or Offline) and why.
Agilent xCELLigence RTCA eSight
🔴 Offline
Out of service; awaiting further information and service action.
Last QC
—
—
Last Maintenance
2026-08-11
2026-08-11
Deltavision OMX
🔴 Offline
Out of service. Only 405 nm and 647 nm TIRF are reported working; the system is not available for general use.
Last QC
—
—
Last Maintenance
2026-08-11
2026-08-11
Leica STELLARIS 8 FALCON FLIM
🟡 Warning
Limited service: avoid 405 nm excitation-dependent experiments. Other wavelengths remain functional.
Last QC
—
—
Last Maintenance
2026-06-16
2026-06-16
Zeiss LSM 880 with AiryScan
🟡 Warning
Limited service: images remain usable, especially with unidirectional scanning, but the scan-field test failed and some bidirectional scan speeds show the issue. A VM800 startup communication error also remains open. The 40x multi-immersion objective's correction collar was repaired and the objective returned to normal operation.
Last QC
2026-05-13
2026-05-13
Last Maintenance
2026-05-13
2026-05-13
Zeiss TIRF
🟡 Warning
Limited to widefield imaging only
Last QC
—
—
Last Maintenance
2026-02-22
2026-02-22