Skip to content
← Advanced Imaging Core Facility at Turku Bioscience Centre

Instrument status

This page lists instruments currently marked Warning or Offline, together with the recorded reason and latest QC and maintenance dates.

Deltavision OMX
GE Healthcare · OMX V4
🔴 Offline Out of service. Only 405 nm and 647 nm TIRF are reported working; the system is not available for general use.
Last QC
—
Last maintenance
2026-08-11
Leica STELLARIS 8 FALCON FLIM
Leica Microsystems · STELLARIS 8 FALCON
🟡 Warning Limited service: avoid 405 nm excitation-dependent experiments. Other wavelengths remain functional.
Last QC
—
Last maintenance
2026-06-16
Zeiss LSM 880 with AiryScan
Zeiss · LSM 880 with AiryScan
🟡 Warning Limited service: images remain usable, especially with unidirectional scanning, but the scan-field test failed and some bidirectional scan speeds show the issue. A VM800 startup communication error also remains open. The 40x multi-immersion objective's correction collar was repaired and the objective returned to normal operation.
Last QC
2026-05-13
Last maintenance
2026-05-13
Zeiss TIRF
Zeiss · TIRF
🟡 Warning Limited to widefield imaging only
Last QC
—
Last maintenance
2026-02-22