System Health

Work in progress

This dashboard is a work in progress. Please independently verify critical information, as not all features are working yet.

System Health Overview

This page lists instruments currently in a degraded state (Warning or Offline) and why.

Agilent xCELLigence RTCA eSight
Agilent · xCELLigence RTCA eSight
🔴 Offline Out of service; awaiting further information and service action.
Last QC
Last Maintenance
2026-08-11
Deltavision OMX
GE Healthcare · OMX V4
🔴 Offline Out of service. Only 405 nm and 647 nm TIRF are reported working; the system is not available for general use.
Last QC
Last Maintenance
2026-08-11
Leica STELLARIS 8 FALCON FLIM
Leica Microsystems · STELLARIS 8 FALCON
🟡 Warning Limited service: avoid 405 nm excitation-dependent experiments. Other wavelengths remain functional.
Last QC
Last Maintenance
2026-06-16
Zeiss LSM 880 with AiryScan
Zeiss · LSM 880 with AiryScan
🟡 Warning Limited service: images remain usable, especially with unidirectional scanning, but the scan-field test failed and some bidirectional scan speeds show the issue. A VM800 startup communication error also remains open. The 40x multi-immersion objective's correction collar was repaired and the objective returned to normal operation.
Last QC
2026-05-13
Last Maintenance
2026-05-13
Zeiss TIRF
Zeiss · TIRF
🟡 Warning Limited to widefield imaging only
Last QC
Last Maintenance
2026-02-22