Skip to content

📝 Event: 2019-03-12_preventative_maintenance

Work in progress

This dashboard is a work in progress. Please independently verify critical information, as not all features are working yet.

📊 Report Overview

Event Metadata
Record Type
Maintenance Event
Reason / Suite
Preventative_maintenance
Service Provider
vendor
Instrument
scope-3i-csu-w1-spinning-disk
Maintenance Action
Reason Details:
preventative_maintenance
Action Details:
Preventative Maintenance Visit. CSU-W was upgraded with YOKOGAWA dust-measurements (sponge for cable-optic-tunnel) and internal cleaning. To address EMCCD connection issues, the driver was updated to PVCAM 3.7.5.7, which resolved the failure. 10x objective lens head noted as jammed. Incubator housing clamp remains broken. Kohler, Vector, and FRAP light paths aligned.

📄 Raw YAML Ledger

View 2019-03-12_preventative_maintenance.yaml
record_type: maintenance_event
microscope: scope-3i-csu-w1-spinning-disk
maintenance_id: maint-2019-03-12-pm
started_utc: '2019-03-12T09:00:00Z'
completed_utc: '2019-03-12T17:00:00Z'
status: completed
reason: preventative_maintenance
action: software_update
action_details: 'Preventative Maintenance Visit. CSU-W was upgraded with YOKOGAWA
  dust-measurements (sponge for cable-optic-tunnel) and internal cleaning.  To address
  EMCCD connection issues, the driver was updated to PVCAM 3.7.5.7, which resolved
  the failure.  10x objective lens head noted as jammed. Incubator housing clamp remains
  broken.  Kohler, Vector, and FRAP light paths aligned.

  '
service_provider: vendor
company: 3i (Intelligent Imaging Innovations)
contact: 3i Field Service Engineer
status_notes: Fully functional