2019-03-12_preventative_maintenanceMaintenance action
Reason details:
scheduled
scheduled
Action details:
Preventative Maintenance Visit. CSU-W was upgraded with YOKOGAWA dust-measurements (sponge for cable-optic-tunnel) and internal cleaning. To address EMCCD connection issues, the driver was updated to PVCAM 3.7.5.7, which resolved the failure. 10x objective lens head noted as jammed. Incubator housing clamp remains broken. Kohler, Vector, and FRAP light paths aligned.
Preventative Maintenance Visit. CSU-W was upgraded with YOKOGAWA dust-measurements (sponge for cable-optic-tunnel) and internal cleaning. To address EMCCD connection issues, the driver was updated to PVCAM 3.7.5.7, which resolved the failure. 10x objective lens head noted as jammed. Incubator housing clamp remains broken. Kohler, Vector, and FRAP light paths aligned.
Raw YAML record
View 2019-03-12_preventative_maintenance.yaml
record_type: maintenance_event
microscope: scope-3i-csu-w1-spinning-disk
maintenance_id: maint-2019-03-12-pm
started_utc: '2019-03-12T09:00:00Z'
completed_utc: '2019-03-12T17:00:00Z'
status: completed
reason: scheduled
action: update
action_details: 'Preventative Maintenance Visit. CSU-W was upgraded with YOKOGAWA
dust-measurements (sponge for cable-optic-tunnel) and internal cleaning. To address
EMCCD connection issues, the driver was updated to PVCAM 3.7.5.7, which resolved
the failure. 10x objective lens head noted as jammed. Incubator housing clamp remains
broken. Kohler, Vector, and FRAP light paths aligned.
'
service_provider: vendor
company: 3i (Intelligent Imaging Innovations)
contact: 3i Field Service Engineer
status_notes: Fully functional