Maintenance event · 2026-03-27

3i CSU-W1 Spinning Disk

Event overview

Event metadata
Record type
Maintenance event
Reason / suite
Problem
Service provider
Vendor
Record ID
2026-03-27_objective_change_issue_fixed
Maintenance action
Reason details:
Objective change and z-drive operation produced abnormal noise and intermittent jamming.
Action details:
Inspected the objective and z-drive system, found and replaced one cable with a bad connection, re-seated the PFS mirror that had shifted out of position and caused mechanical jamming, and cleaned accumulated dirt from the affected mechanism.
Status after event:
Limited Service

Additional notes

Maintenance report: objective change issue is fixed. The system was checked end-to-end. A faulty cable connection was identified and replaced. The issue was reproducible until the root cause was confirmed as a displaced PFS mirror that intermittently jammed the mechanism. After re-seating the mirror and cleaning dirt from the assembly, movement returned to smooth operation. The microscope remains in limited status due to separate unresolved issues: increasing laser startup delay, occasional camera errors, and a SlideBook montage definition problem in the Y-direction.

Raw YAML record

View 2026-03-27_objective_change_issue_fixed.yaml
schema_version: 1
record_type: maintenance_event
maintenance_id: maint_scope-3i-csu-w1-spinning-disk_20260327_objective-change-fix
microscope: scope-3i-csu-w1-spinning-disk
service_provider: vendor
company: ''
contact: ''
reference: ''
started_utc: '2026-03-27T08:30:00Z'
ended_utc: '2026-03-27T13:00:00Z'
system_down_since_utc: '2026-03-27T08:30:00Z'
reason: problem
reason_details: Objective change and z-drive operation produced abnormal noise and
  intermittent jamming.
action: repair
action_details: Inspected the objective and z-drive system, found and replaced one
  cable with a bad connection, re-seated the PFS mirror that had shifted out of position
  and caused mechanical jamming, and cleaned accumulated dirt from the affected mechanism.
microscope_status_before: limited
microscope_status_after: limited
downtime_hours: 4.5
related_qc: []
related_maintenance:
- maint-2025-04-02-zdrive
attachments: []
tags:
- objective
- z-drive
- pfs
- cleaning
followup: Objective/z-drive issue resolved, but system remains limited pending separate
  diagnostics for laser startup delay, intermittent camera errors, and SlideBook Y-direction
  montage bug.
next_due_date: ''
notes: 'Maintenance report: objective change issue is fixed. The system was checked
  end-to-end. A faulty cable connection was identified and replaced. The issue was
  reproducible until the root cause was confirmed as a displaced PFS mirror that intermittently
  jammed the mechanism. After re-seating the mirror and cleaning dirt from the assembly,
  movement returned to smooth operation. The microscope remains in limited status
  due to separate unresolved issues: increasing laser startup delay, occasional camera
  errors, and a SlideBook montage definition problem in the Y-direction.'
status_notes: Objective issue fixed, but laser startup delay, camera errors, and Y-montage
  bug remain