📝 Event: 2026-02-24_installation_acceptance_and_laser_baseline

Work in progress

This dashboard is a work in progress. Please independently verify critical information, as not all features are working yet.

📊 Report Overview

Event Metadata
Record Type
Maintenance Event
Reason / Suite
Install
Service Provider
vendor
Instrument
scope-andor-bc43
Maintenance Action
Reason Details:
Post-installation setup verification and acceptance testing after BC43 delivery.
Action Details:
Ensured correct setup and Fusion/Windows-related settings. Tested imaging workflow from sample-holder and incubation chamber setup/calibration through Montage, Z-Scan, Multichannel Transmitted Light, and Joystick operation using provided sample slides. Verified integrity of objectives and optical pathway. Completed setup and calibration of the Okolab incubation system. Measured laser output using calibrated laser slides: 405 nm = 18 mW, 488 nm = 12 mW, 561 nm = 13 mW, 638 nm = 16 mW.
Status After Fix:
IN_SERVICE

Additional Notes

Installation engineer completed commissioning checks and confirmed instrument readiness for routine operation.

📄 Raw YAML Ledger

View 2026-02-24_installation_acceptance_and_laser_baseline.yaml
schema_version: 1
record_type: maintenance_event
maintenance_id: maint_scope-andor-bc43_20260224_installation_acceptance
microscope: scope-andor-bc43
service_provider: vendor
company: Andor
reference: BC43 installation and acceptance visit
started_utc: '2026-02-24T08:30:00Z'
ended_utc: '2026-02-24T12:30:00Z'
reason: install
reason_details: 'Post-installation setup verification and acceptance testing after
  BC43 delivery.

  '
action: calibrate
action_details: 'Ensured correct setup and Fusion/Windows-related settings. Tested
  imaging workflow from sample-holder and incubation chamber setup/calibration through
  Montage, Z-Scan, Multichannel Transmitted Light, and Joystick operation using provided
  sample slides. Verified integrity of objectives and optical pathway. Completed setup
  and calibration of the Okolab incubation system. Measured laser output using calibrated
  laser slides: 405 nm = 18 mW, 488 nm = 12 mW, 561 nm = 13 mW, 638 nm = 16 mW.

  '
microscope_status_before: out_of_service
microscope_status_after: in_service
downtime_hours: 4.0
related_qc:
- qc_scope-andor-bc43_20260224T1100Z_laser_baseline
attachments: []
tags:
- installation
- acceptance
- calibration
- laser-power
- okolab
followup: 'Use calibrated laser slides for routine power baseline checks and log future
  values in QC sessions.

  '
next_due_date: '2026-03-24'
notes: 'Installation engineer completed commissioning checks and confirmed instrument
  readiness for routine operation.

  '
contact: Installation engineer
status_notes: Fully functional