📝 Event: 2026-02-24_installation_laser_baseline_qc
Work in progress
This dashboard is a work in progress. Please independently verify critical information, as not all features are working yet.
📊 Report Overview
Event Metadata
Record Type
Qc Session
Qc Session
Reason / Suite
Post_installation_baseline
Post_installation_baseline
Operator
Installation engineer
Installation engineer
Instrument
scope-andor-bc43QC Summary & Evaluation
Summary: Baseline laser power check completed during BC43 installation acceptance.
🔬 Tests Performed
| Test Type | Details |
|---|---|
| Laser Power | Single-point baseline measurement for each installed laser line. |
📏 Computed Metrics
| Metric ID | Value |
|---|---|
laser.at_obj.405.100pct.power_mw |
18.0 |
laser.at_obj.488.100pct.power_mw |
12.0 |
laser.at_obj.561.100pct.power_mw |
13.0 |
laser.at_obj.638.100pct.power_mw |
16.0 |
Additional Notes
Laser power measured with calibrated laser slides provided with the instrument package. Values captured as initial reference for future trending.
📄 Raw YAML Ledger
View 2026-02-24_installation_laser_baseline_qc.yaml
schema_version: 1
record_type: qc_session
qc_id: qc_scope-andor-bc43_20260224T1100Z_laser_baseline_v1
microscope: scope-andor-bc43
started_utc: '2026-02-24T11:00:00Z'
ended_utc: '2026-02-24T11:20:00Z'
reason: post_installation_baseline
summary: Baseline laser power check completed during BC43 installation acceptance.
notes: 'Laser power measured with calibrated laser slides provided with the instrument
package. Values captured as initial reference for future trending.
'
performed:
- qc_type: laser_power
details: Single-point baseline measurement for each installed laser line.
artifacts: []
inputs_human:
- metric_id: visual.optical_path_integrity
value: verified
unit: ''
details: Objectives and optical pathway integrity checked during installation acceptance.
metric_class: signal_to_noise
laser_inputs_human:
measurement_position: at_objective
measurement_position_details: Measured using calibrated laser slides supplied with
the facility.
power_meter:
model: calibrated laser slides
serial: ''
calibration_due: ''
integration_time_s: 1.0
notes: Installation acceptance baseline.
linearity_series:
- laser: '405'
setpoint_units: percent
power_units: mW
points:
- setpoint: 100
power: 18.0
details: Baseline measured during installation acceptance.
- laser: '488'
setpoint_units: percent
power_units: mW
points:
- setpoint: 100
power: 12.0
details: Baseline measured during installation acceptance.
- laser: '561'
setpoint_units: percent
power_units: mW
points:
- setpoint: 100
power: 13.0
details: Baseline measured during installation acceptance.
- laser: '638'
setpoint_units: percent
power_units: mW
points:
- setpoint: 100
power: 16.0
details: Baseline measured during installation acceptance.
stability_series: []
computed_provenance:
pipeline: manual_entry
version: '1'
run_utc: '2026-02-24T11:30:00Z'
metrics_computed:
- metric_id: laser.at_obj.405.100pct.power_mw
value: 18.0
unit: mW
details: Baseline max power at objective measured during installation.
metric_class: laser_power
- metric_id: laser.at_obj.488.100pct.power_mw
value: 12.0
unit: mW
details: Baseline max power at objective measured during installation.
metric_class: laser_power
- metric_id: laser.at_obj.561.100pct.power_mw
value: 13.0
unit: mW
details: Baseline max power at objective measured during installation.
metric_class: laser_power
- metric_id: laser.at_obj.638.100pct.power_mw
value: 16.0
unit: mW
details: Baseline max power at objective measured during installation.
metric_class: laser_power
artifacts: []
performed_by: Installation engineer