📝 Event: 2026-02-24_installation_laser_baseline_qc

Work in progress

This dashboard is a work in progress. Please independently verify critical information, as not all features are working yet.

📊 Report Overview

Event Metadata
Record Type
Qc Session
Reason / Suite
Post_installation_baseline
Operator
Installation engineer
Instrument
scope-andor-bc43
QC Summary & Evaluation

Summary: Baseline laser power check completed during BC43 installation acceptance.

🔬 Tests Performed

Test Type Details
Laser Power Single-point baseline measurement for each installed laser line.

📏 Computed Metrics

Metric ID Value
laser.at_obj.405.100pct.power_mw 18.0
laser.at_obj.488.100pct.power_mw 12.0
laser.at_obj.561.100pct.power_mw 13.0
laser.at_obj.638.100pct.power_mw 16.0

Additional Notes

Laser power measured with calibrated laser slides provided with the instrument package. Values captured as initial reference for future trending.

📄 Raw YAML Ledger

View 2026-02-24_installation_laser_baseline_qc.yaml
schema_version: 1
record_type: qc_session
qc_id: qc_scope-andor-bc43_20260224T1100Z_laser_baseline_v1
microscope: scope-andor-bc43
started_utc: '2026-02-24T11:00:00Z'
ended_utc: '2026-02-24T11:20:00Z'
reason: post_installation_baseline
summary: Baseline laser power check completed during BC43 installation acceptance.
notes: 'Laser power measured with calibrated laser slides provided with the instrument
  package. Values captured as initial reference for future trending.

  '
performed:
- qc_type: laser_power
  details: Single-point baseline measurement for each installed laser line.
  artifacts: []
inputs_human:
- metric_id: visual.optical_path_integrity
  value: verified
  unit: ''
  details: Objectives and optical pathway integrity checked during installation acceptance.
  metric_class: signal_to_noise
laser_inputs_human:
  measurement_position: at_objective
  measurement_position_details: Measured using calibrated laser slides supplied with
    the facility.
  power_meter:
    model: calibrated laser slides
    serial: ''
    calibration_due: ''
    integration_time_s: 1.0
    notes: Installation acceptance baseline.
  linearity_series:
  - laser: '405'
    setpoint_units: percent
    power_units: mW
    points:
    - setpoint: 100
      power: 18.0
    details: Baseline measured during installation acceptance.
  - laser: '488'
    setpoint_units: percent
    power_units: mW
    points:
    - setpoint: 100
      power: 12.0
    details: Baseline measured during installation acceptance.
  - laser: '561'
    setpoint_units: percent
    power_units: mW
    points:
    - setpoint: 100
      power: 13.0
    details: Baseline measured during installation acceptance.
  - laser: '638'
    setpoint_units: percent
    power_units: mW
    points:
    - setpoint: 100
      power: 16.0
    details: Baseline measured during installation acceptance.
  stability_series: []
computed_provenance:
  pipeline: manual_entry
  version: '1'
  run_utc: '2026-02-24T11:30:00Z'
metrics_computed:
- metric_id: laser.at_obj.405.100pct.power_mw
  value: 18.0
  unit: mW
  details: Baseline max power at objective measured during installation.
  metric_class: laser_power
- metric_id: laser.at_obj.488.100pct.power_mw
  value: 12.0
  unit: mW
  details: Baseline max power at objective measured during installation.
  metric_class: laser_power
- metric_id: laser.at_obj.561.100pct.power_mw
  value: 13.0
  unit: mW
  details: Baseline max power at objective measured during installation.
  metric_class: laser_power
- metric_id: laser.at_obj.638.100pct.power_mw
  value: 16.0
  unit: mW
  details: Baseline max power at objective measured during installation.
  metric_class: laser_power
artifacts: []
performed_by: Installation engineer