📝 Event: 2025-12-22_wll_replacement

Work in progress

This dashboard is a work in progress. Please independently verify critical information, as not all features are working yet.

📊 Report Overview

Event Metadata
Record Type
Maintenance Event
Reason / Suite
Problem
Service Provider
vendor
Instrument
scope-leica-stellaris-8-falcon-flim-resonant
Maintenance Action
Reason Details:
Vendor service visit for an ongoing White Light Laser (WLL) issue on the Leica STELLARIS FALCON system. The service report describes the complaint as "WLL replacement" and notes that the issue had been ongoing in Finland for several months, with a related earlier case referenced as CS 397993.
Action Details:
Leica Microsystems replaced/integrated a new White Light Laser into the system. The service report states: "New WLL has been integrated to the system. All laser lines of the WLL are totally in spec."
Status After Fix:
IN_SERVICE

Additional Notes

Source document: Leica Microsystems Technical Report SA-223726_V1.pdf. Reported asset details: Equipment STELLARIS FALCON - 13334341, serial number 8300000681, product 158305100, product family X01. The report lists operations CALL4 and TRVL4, signed by Alexander Ponomarev on 2025-12-23 06:23. The report does not provide measured laser powers, so this maintenance event records the vendor statement only and does not create structured laser-power QC values. No specific follow-up action was stated in the service report; continue routine laser/QC monitoring and record measured WLL line powers in a QC session if local measurements are performed.

📄 Raw YAML Ledger

View 2025-12-22_wll_replacement.yaml
schema_version: 1
record_type: maintenance_event
maintenance_id: maint_scope-leica-stellaris-8-falcon-flim-resonant_20251222_wll_replacement
microscope: scope-leica-stellaris-8-falcon-flim-resonant

service_provider: vendor
company: Leica Microsystems
contact: Leica Microsystems field service engineer
reference: 'Case CS-00401027; Work order WO-00279399; Appointment SA-223726; SAP service order 002000646116; contract 43295213'

started_utc: '2025-12-22T08:00:00Z'
ended_utc: '2025-12-23T09:00:00Z'

reason: problem
reason_details: >-
  Vendor service visit for an ongoing White Light Laser (WLL) issue on the Leica
  STELLARIS FALCON system. The service report describes the complaint as "WLL
  replacement" and notes that the issue had been ongoing in Finland for several
  months, with a related earlier case referenced as CS 397993.

action: replace
action_details: >-
  Leica Microsystems replaced/integrated a new White Light Laser into the system.
  The service report states: "New WLL has been integrated to the system. All
  laser lines of the WLL are totally in spec."

microscope_status_before: limited
microscope_status_after: in_service
status_notes: >-
  WLL replacement completed by Leica Microsystems. The service report states that
  All WLL laser lines were in specification after integration, and the system
  was fully tested and passed all applicable acceptance criteria.

downtime_hours: 25.0

attachments:
  - maintenance/attachments/scope-leica-stellaris-8-falcon-flim-resonant/2025/SA-223726_V1.pdf

tags:
  - WLL
  - white_light_laser
  - laser
  - vendor_service
  - warranty_extension

notes: >-
  Source document: Leica Microsystems Technical Report SA-223726_V1.pdf. Reported
  asset details: Equipment STELLARIS FALCON - 13334341, serial number 8300000681,
  product 158305100, product family X01. The report lists operations CALL4 and
  TRVL4, signed by Alexander Ponomarev on 2025-12-23 06:23. The report does not
  provide measured laser powers, so this maintenance event records the vendor
  statement only and does not create structured laser-power QC values. No specific
  follow-up action was stated in the service report; continue routine laser/QC
  monitoring and record measured WLL line powers in a QC session if local
  measurements are performed.