📝 Event: 2025-02-11_preventive_maintenance
Work in progress
This dashboard is a work in progress. Please independently verify critical information, as not all features are working yet.
📊 Report Overview
Event Metadata
Record Type
Maintenance Event
Maintenance Event
Reason / Suite
Scheduled
Scheduled
Service Provider
vendor
vendor
Instrument
scope-testx1Maintenance Action
Reason Details:
Annual preventive maintenance visit after high usage in Q4. Users had noticed mild focus drift during long 60x live-cell runs.
Annual preventive maintenance visit after high usage in Q4. Users had noticed mild focus drift during long 60x live-cell runs.
Action Details:
Completed vendor PM checklist, cleaned objective turret contacts, checked stage cable strain relief, and adjusted incubation chamber fan profile to reduce vibration transfer.
Completed vendor PM checklist, cleaned objective turret contacts, checked stage cable strain relief, and adjusted incubation chamber fan profile to reduce vibration transfer.
Status After Fix:
IN_SERVICE
Additional Notes
Drift reduced after fan profile adjustment. No component replacement needed.
📄 Raw YAML Ledger
View 2025-02-11_preventive_maintenance.yaml
schema_version: 1
record_type: maintenance_event
maintenance_id: maint_scope-testx1_20250211_preventive_maintenance
microscope: scope-testx1
service_provider: vendor
company: Nordic Imaging Service Oy
reference: NIS-2025-0211-044
started_utc: '2025-02-11T07:45:00Z'
ended_utc: '2025-02-11T11:20:00Z'
reason: scheduled
reason_details: 'Annual preventive maintenance visit after high usage in Q4. Users
had noticed mild focus drift during long 60x live-cell runs.
'
action: service
action_details: 'Completed vendor PM checklist, cleaned objective turret contacts,
checked stage cable strain relief, and adjusted incubation chamber fan profile to
reduce vibration transfer.
'
microscope_status_after: in_service
downtime_hours: 3.6
related_qc:
- qc_scope-testx1_20250210T0910Z_psf_v1
- qc_scope-testx1_20250212T0835Z_psf_v1
attachments:
- maintenance/events/scope-testx1/2025/attachments/2025-02-11_pm_report.pdf
tags:
- scheduled
- focus
- vibration
- vendor
followup: 'Re-check autofocus repeatability after two weeks of incubated imaging.
'
next_due_date: '2026-02-01'
notes: 'Drift reduced after fan profile adjustment. No component replacement needed.
'
contact: 'Field Engineer: Laura Niemi (laura.niemi@nordicimaging.fi)'
status_notes: Fully functional