📝 Event: 2025-04-02_preventive_maintenance

Work in progress

This dashboard is a work in progress. Please independently verify critical information, as not all features are working yet.

📊 Report Overview

Event Metadata
Record Type
Maintenance Event
Reason / Suite
Scheduled
Service Provider
vendor
Instrument
scope-zeiss-lsm-880-with-airyscan
Maintenance Action
Reason Details:
Annual preventive maintenance visit for the Zeiss LSM 880 / Axio Observer 7. The visit was triggered by the routine vendor PM schedule and included follow-up on reported stage movement irregularities and reduced argon laser performance.
Action Details:
Vendor preventive maintenance was performed according to the Zeiss protocol. Checks passed for environment, cabling, system startup, anti-vibration table, air filters, surge protectors, ECU voltages, electrical and laser safety, PC ventilation, microscope general function, liquid-cooling related items, detector/calibration routines, beam profile, AOTF, scan field, pinhole/MBS, and Airyscan alignment. MBS filter wheel axles were cleaned and lubricated. Maincontroller battery was checked at 3.03 V and not replaced. Manual stage control and tile-scan test were completed, but stage motion was reported as occasionally not smooth. The argon laser output was low and the engineer documented it as end of life, so some maintenance tasks including the SNR test could not be completed and were deferred until after argon laser replacement.
Status After Fix:
LIMITED

Additional Notes

Timeline anchored to the field service report completion date (2025-04-02). The signed maintenance checklist header shows 2025-03-24, so the repository timestamp was normalized to 2025-04-02T12:00:00Z because the source documents do not provide an authoritative visit start time. Final remarks state that the instrument is not fully operational: argon lines 488/514 are low and out of spec, and the stage may have a mechanical issue. Additional documented context: ZEN 2.3 SP1 FP3 HF 29 FP3, CanCheck 13.0.19.1, RT PC MICO HS, user PC RAM 192 GB, Windows 7 64-bit. Recommended follow-up: Replace the argon laser, then repeat the pending maintenance tasks and detector SNR test. Review stage mechanics with PI and decide between service/maintenance by PI or replacement with a new stage.

📄 Raw YAML Ledger

View 2025-04-02_preventive_maintenance.yaml
schema_version: 1
record_type: maintenance_event
maintenance_id: maint_scope-zeiss-lsm-880-with-airyscan_20250402_preventive_maintenance
microscope: scope-zeiss-lsm-880-with-airyscan
service_provider: vendor
company: Carl Zeiss Oy
contact: field service engineer
reference: Service order 8583415960 / 10; external reference CCC-001568
started_utc: '2025-04-02T12:00:00Z'
reason: scheduled
reason_details: Annual preventive maintenance visit for the Zeiss LSM 880 / Axio Observer
  7. The visit was triggered by the routine vendor PM schedule and included follow-up
  on reported stage movement irregularities and reduced argon laser performance.
action: service
action_details: Vendor preventive maintenance was performed according to the Zeiss
  protocol. Checks passed for environment, cabling, system startup, anti-vibration
  table, air filters, surge protectors, ECU voltages, electrical and laser safety,
  PC ventilation, microscope general function, liquid-cooling related items, detector/calibration
  routines, beam profile, AOTF, scan field, pinhole/MBS, and Airyscan alignment. MBS
  filter wheel axles were cleaned and lubricated. Maincontroller battery was checked
  at 3.03 V and not replaced. Manual stage control and tile-scan test were completed,
  but stage motion was reported as occasionally not smooth. The argon laser output
  was low and the engineer documented it as end of life, so some maintenance tasks
  including the SNR test could not be completed and were deferred until after argon
  laser replacement.
microscope_status_after: limited
status_notes: Usable, but limited by end-of-life argon output and occasional stage
  movement hiccups.
related_qc:
- qc_scope-zeiss-lsm-880-with-airyscan_20250402T1200Z_post_service_laser_stage_v1
attachments:
- maintenance/events/scope-zeiss-lsm-880-with-airyscan/2025/attachments/Service_report_8583415960-10_2025-04-02_13-55.pdf
- maintenance/events/scope-zeiss-lsm-880-with-airyscan/2025/attachments/Signed_MaintenanceChecklist_LSM880_2018-09-25.pdf
tags:
- scheduled
- vendor
- argon-laser
- stage
- airyscan
notes: 'Timeline anchored to the field service report completion date (2025-04-02).
  The signed maintenance checklist header shows 2025-03-24, so the repository timestamp
  was normalized to 2025-04-02T12:00:00Z because the source documents do not provide
  an authoritative visit start time. Final remarks state that the instrument is not
  fully operational: argon lines 488/514 are low and out of spec, and the stage may
  have a mechanical issue. Additional documented context: ZEN 2.3 SP1 FP3 HF 29 FP3,
  CanCheck 13.0.19.1, RT PC MICO HS, user PC RAM 192 GB, Windows 7 64-bit. Recommended
  follow-up: Replace the argon laser, then repeat the pending maintenance tasks and
  detector SNR test. Review stage mechanics with PI and decide between service/maintenance
  by PI or replacement with a new stage.'