📝 Event: 2026-05-13_preventive_maintenance

Work in progress

This dashboard is a work in progress. Please independently verify critical information, as not all features are working yet.

📊 Report Overview

Event Metadata
Record Type
Maintenance Event
Reason / Suite
Scheduled
Service Provider
vendor
Instrument
scope-zeiss-lsm-880-with-airyscan
Maintenance Action
Reason Details:
Annual vendor preventive maintenance for the Zeiss LSM 880 AS Fast. The official checklist documents routine system, safety, cooling, laser, calibration, detector, stage, focus, and application-level checks. The same vendor visit also included corrective follow-up for the previously reported broken correction collar on the 40x multi-immersion objective.
Action Details:
Vendor preventive maintenance was completed. Environment, cabling, system condition, anti-vibration table, fans and filters, surge protection, ECU voltage lines, electrical safety, laser safety, microscope function, external cooling, equipotential bonding, detector function and gain response, InVis collimation, QUASAR spectral calibration, V-laser attenuator, laser beam profiles, AOTF settings, resolution, stepper-motor test, laser linearization, pinhole and MBS calibrations, detector offset and amplification, PMT linearization, detector SNR checks, XY stage control, Z focus, Z stack, visible-channel application tests, and Airyscan tests were recorded in order. PC ventilation was cleaned; the cooling liquid was replaced; MBS filter-wheel axles were cleaned and lubricated; CanCheck was updated to 14.0.8.1; and Airyscan alignment was checked and fine-tuned. The scan-field test failed, indicating scanner wear, and the vendor recommended replacement of the scanners and scanner drive boards. A Can29_E1 VM800 communication error was also observed when ZEN started. During the same visit, the field engineer repaired the correction collar on the LD LCI Plan APOCHROMAT 40x/1.2 Imm Korr DIC (UV)VIS-IR objective (product code 420862-9970, serial 4932000043) and verified objective operation. The repair was confirmed on 2026-05-14.
Status After Fix:
LIMITED

Additional Notes

The source provides a service date but no authoritative start time, so started_utc is normalized to 00:00 UTC on 2026-05-13. Instrument serial 2850000143. The checklist marks the instrument as not fully operational. The signed source report is not stored in this public repository because it contains personal contact and signature data; all ledger text is anonymized. This consolidated event closes the 2026-05-04 correction-collar issue for hardware.objectives[id=40x_multi]. It does not close the independent scanner or VM800 communication issues documented here.

📄 Raw YAML Ledger

View 2026-05-13_preventive_maintenance.yaml
schema_version: 1
record_type: maintenance_event
maintenance_id: maint_scope-zeiss-lsm-880-with-airyscan_20260513_preventive_maintenance
microscope: scope-zeiss-lsm-880-with-airyscan

service_provider: vendor
company: Carl Zeiss Oy
contact: field service engineer
reference: >-
  Worksheet 8583910132; field engineer objective-repair confirmation dated
  2026-05-14

started_utc: '2026-05-13T00:00:00Z'
ended_utc: '2026-05-14T09:00:00Z'

reason: scheduled
reason_details: >-
  Annual vendor preventive maintenance for the Zeiss LSM 880 AS Fast. The official
  checklist documents routine system, safety, cooling, laser, calibration, detector,
  stage, focus, and application-level checks. The same vendor visit also included
  corrective follow-up for the previously reported broken correction collar on the
  40x multi-immersion objective.

action: service
action_details: >-
  Vendor preventive maintenance was completed. Environment, cabling, system condition,
  anti-vibration table, fans and filters, surge protection, ECU voltage lines,
  electrical safety, laser safety, microscope function, external cooling, equipotential
  bonding, detector function and gain response, InVis collimation, QUASAR spectral
  calibration, V-laser attenuator, laser beam profiles, AOTF settings, resolution,
  stepper-motor test, laser linearization, pinhole and MBS calibrations, detector offset
  and amplification, PMT linearization, detector SNR checks, XY stage control, Z focus,
  Z stack, visible-channel application tests, and Airyscan tests were recorded in order.
  PC ventilation was cleaned; the cooling liquid was replaced; MBS filter-wheel axles
  were cleaned and lubricated; CanCheck was updated to 14.0.8.1; and Airyscan alignment
  was checked and fine-tuned. The scan-field test failed, indicating scanner wear, and
  the vendor recommended replacement of the scanners and scanner drive boards. A
  Can29_E1 VM800 communication error was also observed when ZEN started.
  During the same visit, the field engineer repaired the correction collar on the
  LD LCI Plan APOCHROMAT 40x/1.2 Imm Korr DIC (UV)VIS-IR objective (product code
  420862-9970, serial 4932000043) and verified objective operation. The repair was
  confirmed on 2026-05-14.

microscope_status_before: limited
microscope_status_after: limited
status_notes: >-
  Limited service: images remain usable, especially with unidirectional scanning, but
  the scan-field test failed and some bidirectional scan speeds show the issue. A VM800
  startup communication error also remains open. The 40x multi-immersion objective's
  correction collar was repaired and the objective returned to normal operation.

related_qc:
  - qc_scope-zeiss-lsm-880-with-airyscan_20260513T0000Z_vendor_pm_v1

related_maintenance:
  - maint_scope-zeiss-lsm-880-with-airyscan_20260504_40x_objective_correction_collar_issue

tags:
  - scheduled
  - vendor
  - scanner
  - scan-field
  - vm800
  - airyscan
  - objective
  - correction_collar
  - repair
  - resolved

followup: >-
  Replace the scanners and scanner drive boards to resolve the failed scan-field test.
  Troubleshoot the VM800 communication error and the MICO HS LVDS interface during the
  next preventive-maintenance visit.

notes: >-
  The source provides a service date but no authoritative start time, so started_utc is
  normalized to 00:00 UTC on 2026-05-13. Instrument serial 2850000143. The checklist
  marks the instrument as not fully operational. The signed source report is not stored
  in this public repository because it contains personal contact and signature data;
  all ledger text is anonymized. This consolidated event closes the 2026-05-04
  correction-collar issue for hardware.objectives[id=40x_multi]. It does not close the
  independent scanner or VM800 communication issues documented here.