馃搳 Report Overview
Qc Session
Preventive_maintenance
Zeiss field service engineer
scope-zeiss-lsm-880-with-airyscanSummary: Vendor preventive-maintenance QC recorded laser power, detector SNR, stage position error, optical alignment, focus, and application-level checks. The scan-field test failed while the remaining recorded QC checks were in order.
馃敩 Tests Performed
| Test Type | Details |
|---|---|
| Laser Power | Vendor single-point laser-power measurements at the fiber output and objective plane, including argon operating current and lifetime counters. |
| Stage | Manual XY-stage function and 3x3 TileScanRotation.lvb position-error check at a 512-pixel frame size. |
| Alignment | Detector, collimator, spectral, beam-profile, AOTF, resolution, scan-field, pinhole, MBS, PMT, and Airyscan alignment/calibration checks. |
| Psf | Visible-channel application checks included beads and a PSF Z scan using the Plan-Apo 63x oil objective; both were recorded in order. |
馃敠 Laser Measurement Context
| Field | Value |
|---|---|
| Measurement position | at objective |
| Position details | Objective-plane values were measured at 100% output using ZEN Spot scan and a Plan-Neofluar 10x test objective. Fiber-output measurements are preserved in inputs_human because the schema permits one primary measurement position per QC session. |
馃搹 QC Measurements
| Measurement | Value | Source | Details |
|---|---|---|---|
405 nm laser power at objective (100%)laser.at_obj.405.100pct.power_mw |
3.83 mW | Laser input | Objective-plane measurement; checklist result in order. |
458 nm laser power at objective (100%)laser.at_obj.458.100pct.power_mw |
0.19 mW | Laser input | Objective-plane argon-line measurement; checklist result in order. |
488 nm laser power at objective (100%)laser.at_obj.488.100pct.power_mw |
1.33 mW | Laser input | Objective-plane argon-line measurement; checklist result in order. |
543 nm laser power at objective (100%)laser.at_obj.543.100pct.power_mw |
0.27 mW | Laser input | Objective-plane measurement; checklist result in order. |
633 nm laser power at objective (100%)laser.at_obj.633.100pct.power_mw |
1.09 mW | Laser input | Objective-plane measurement; checklist result in order. |
Laser 路 Fiber 路 Argon 路 Optimal Power mWlaser.fiber.argon.optimal_power_mw |
11.42 mW | Human input | Argon power at the documented optimal setting; checklist result in order. |
Laser 路 Fiber 路 Argon 路 Optimal Tube Current Alaser.fiber.argon.optimal_tube_current_a |
6 A | Human input | Argon tube current at the documented optimal setting. |
Laser 路 Argon 路 Total Operating Hourslaser.argon.total_operating_hours |
17550 h | Human input | Argon total operating-hours counter recorded by the vendor. |
Laser 路 Argon 路 Hours Since Exchangelaser.argon.hours_since_exchange |
1903 h | Human input | Argon usage recorded since the previous exchange. |
Laser 路 Fiber 路 405 路 Power mWlaser.fiber.405.power_mw |
13.55 mW | Human input | 405 nm power measured at the fiber output. |
Laser 路 Fiber 路 543 路 Power mWlaser.fiber.543.power_mw |
0.95 mW | Human input | 543 nm power measured at the fiber output. |
Laser 路 Fiber 路 633 路 Power mWlaser.fiber.633.power_mw |
5.12 mW | Human input | 633 nm power measured at the fiber output. |
Signal To Noise 路 Channel 1 Test Statussignal_to_noise.channel_1_test_status |
pass | Human input | Vendor SNR checklist result recorded in order. |
Signal To Noise 路 Channel 2 Test Statussignal_to_noise.channel_2_test_status |
pass | Human input | Vendor SNR checklist result recorded in order. |
Signal To Noise 路 Channel 3 Test Statussignal_to_noise.channel_3_test_status |
pass | Human input | Vendor SNR checklist result recorded in order. |
Signal To Noise 路 32 Channel Array Test Statussignal_to_noise.32_channel_array_test_status |
pass | Human input | Vendor SNR checklist result recorded in order. |
Signal To Noise 路 Airyscan Test Statussignal_to_noise.airyscan_test_status |
pass | Human input | Vendor SNR checklist result recorded in order; no BiG SNR result was entered. |
Stage 路 Tile Scan Error X pxstage.tile_scan_error_x_px |
2 px | Human input | Error in X before stitching in the 3x3 tile-scan test. |
Stage 路 Tile Scan Error Y pxstage.tile_scan_error_y_px |
0 px | Human input | Error in Y before stitching in the 3x3 tile-scan test. |
Stage 路 Tile Scan Error Total pxstage.tile_scan_error_total_px |
4 px | Human input | Vendor-reported total error; acceptance criterion was less than 6 px. |
Additional Notes
Source date only; started_utc is normalized to 00:00 UTC. Objective-plane laser power was measured at 100% output with a Plan-Neofluar 10x test objective. The checklist records the instrument as not fully operational because of the failed scan-field test and separately notes a VM800 startup communication error. The signed source report is retained outside the public repository because it contains personal data.
馃搫 Raw YAML Ledger
View 2026-05-13_vendor_pm_qc.yaml
schema_version: 1
record_type: qc_session
qc_id: qc_scope-zeiss-lsm-880-with-airyscan_20260513T0000Z_vendor_pm_v1
microscope: scope-zeiss-lsm-880-with-airyscan
started_utc: '2026-05-13T00:00:00Z'
performed_by: Zeiss field service engineer
reason: preventive_maintenance
summary: >-
Vendor preventive-maintenance QC recorded laser power, detector SNR, stage position
error, optical alignment, focus, and application-level checks. The scan-field test
failed while the remaining recorded QC checks were in order.
notes: >-
Source date only; started_utc is normalized to 00:00 UTC. Objective-plane laser power
was measured at 100% output with a Plan-Neofluar 10x test objective. The checklist
records the instrument as not fully operational because of the failed scan-field test
and separately notes a VM800 startup communication error. The signed source report is
retained outside the public repository because it contains personal data.
performed:
- qc_type: laser_power
details: >-
Vendor single-point laser-power measurements at the fiber output and objective
plane, including argon operating current and lifetime counters.
- qc_type: stage
details: >-
Manual XY-stage function and 3x3 TileScanRotation.lvb position-error check at a
512-pixel frame size.
- qc_type: alignment
details: >-
Detector, collimator, spectral, beam-profile, AOTF, resolution, scan-field,
pinhole, MBS, PMT, and Airyscan alignment/calibration checks.
- qc_type: psf
details: >-
Visible-channel application checks included beads and a PSF Z scan using the
Plan-Apo 63x oil objective; both were recorded in order.
inputs_human:
- metric_id: laser.fiber.argon.optimal_power_mw
metric_class: laser_power
value: 11.42
unit: mW
details: Argon power at the documented optimal setting; checklist result in order.
- metric_id: laser.fiber.argon.optimal_tube_current_a
metric_class: laser_power
value: 6.0
unit: A
details: Argon tube current at the documented optimal setting.
- metric_id: laser.argon.total_operating_hours
metric_class: laser_power
value: 17550
unit: h
details: Argon total operating-hours counter recorded by the vendor.
- metric_id: laser.argon.hours_since_exchange
metric_class: laser_power
value: 1903
unit: h
details: Argon usage recorded since the previous exchange.
- metric_id: laser.fiber.405.power_mw
metric_class: laser_power
value: 13.55
unit: mW
details: 405 nm power measured at the fiber output.
- metric_id: laser.fiber.543.power_mw
metric_class: laser_power
value: 0.95
unit: mW
details: 543 nm power measured at the fiber output.
- metric_id: laser.fiber.633.power_mw
metric_class: laser_power
value: 5.12
unit: mW
details: 633 nm power measured at the fiber output.
- metric_id: signal_to_noise.channel_1_test_status
metric_class: signal_to_noise
value: pass
unit: ''
details: Vendor SNR checklist result recorded in order.
- metric_id: signal_to_noise.channel_2_test_status
metric_class: signal_to_noise
value: pass
unit: ''
details: Vendor SNR checklist result recorded in order.
- metric_id: signal_to_noise.channel_3_test_status
metric_class: signal_to_noise
value: pass
unit: ''
details: Vendor SNR checklist result recorded in order.
- metric_id: signal_to_noise.32_channel_array_test_status
metric_class: signal_to_noise
value: pass
unit: ''
details: Vendor SNR checklist result recorded in order.
- metric_id: signal_to_noise.airyscan_test_status
metric_class: signal_to_noise
value: pass
unit: ''
details: Vendor SNR checklist result recorded in order; no BiG SNR result was entered.
- metric_id: stage.tile_scan_error_x_px
metric_class: stage_repeatability
value: 2
unit: px
details: Error in X before stitching in the 3x3 tile-scan test.
- metric_id: stage.tile_scan_error_y_px
metric_class: stage_repeatability
value: 0
unit: px
details: Error in Y before stitching in the 3x3 tile-scan test.
- metric_id: stage.tile_scan_error_total_px
metric_class: stage_repeatability
value: 4
unit: px
details: Vendor-reported total error; acceptance criterion was less than 6 px.
laser_inputs_human:
measurement_position: at_objective
measurement_position_details: >-
Objective-plane values were measured at 100% output using ZEN Spot scan and a
Plan-Neofluar 10x test objective. Fiber-output measurements are preserved in
inputs_human because the schema permits one primary measurement position per QC
session.
single_point_measurements:
- laser: '405'
setpoint: 100
setpoint_units: percent
power: 3.83
power_units: mW
details: Objective-plane measurement; checklist result in order.
- laser: '458'
setpoint: 100
setpoint_units: percent
power: 0.19
power_units: mW
details: Objective-plane argon-line measurement; checklist result in order.
- laser: '488'
setpoint: 100
setpoint_units: percent
power: 1.33
power_units: mW
details: Objective-plane argon-line measurement; checklist result in order.
- laser: '543'
setpoint: 100
setpoint_units: percent
power: 0.27
power_units: mW
details: Objective-plane measurement; checklist result in order.
- laser: '633'
setpoint: 100
setpoint_units: percent
power: 1.09
power_units: mW
details: Objective-plane measurement; checklist result in order.
linearity_series:
- laser: '405'
setpoint_units: percent
power_units: mW
points:
- setpoint: 100
power: 3.83
details: Single-point 100% objective-plane measurement from the vendor checklist.
- laser: '458'
setpoint_units: percent
power_units: mW
points:
- setpoint: 100
power: 0.19
details: Single-point 100% objective-plane argon-line measurement.
- laser: '488'
setpoint_units: percent
power_units: mW
points:
- setpoint: 100
power: 1.33
details: Single-point 100% objective-plane argon-line measurement.
- laser: '543'
setpoint_units: percent
power_units: mW
points:
- setpoint: 100
power: 0.27
details: Single-point 100% objective-plane measurement from the vendor checklist.
- laser: '633'
setpoint_units: percent
power_units: mW
points:
- setpoint: 100
power: 1.09
details: Single-point 100% objective-plane measurement from the vendor checklist.
stability_series: []